On-Site Pipette Calibration Services
This comprehensive service ensures your pipettes are fully functional, clean, and compliant with accuracy standards.
​
2 Levels of Services to choose from
Level 1
: Preventive Maintenance & Calibration
This comprehensive service ensures your pipettes are fully functional, clean, and compliant with accuracy standards.
Included Services:
-
Complete disassembly and cleaning (internal & external)
-
Piston inspection, polishing, and lubrication
-
Replacement of seals, O-rings, and wear parts (as needed)
-
Leak and functional testing
-
Gravimetric calibration at 2 volume points (typically 10%, and 100%)
-
Smudge proof label affixed to the pipette, and Work Order Report generated with PASS/ FAIL Calibration Status.
-
One-year service warranty
$20 / Single Channel



Level 2
Calibration Certificate according to ISO/IES 17025 accreditation
For Clinical labs and GLP/GMP Certification.
This service includes an ISO/IEC 17025:2017 accredited calibration certificate with “As-Left” data based on four gravimetric measurements at low and high volume settings. Key performance metrics—%CV, %Accuracy, and Mean—are automatically calculated and reported.
Measurement uncertainty is accounted for using a 95% confidence level, with an expanded uncertainty (k=2), ensuring that results remain fully within specified tolerance limits.
Customized volume points and specification limits are available upon request.